Research › QEEG, Methods & Mechanisms
Evaluation of commercially available electrodes and gels for recording of slow EEG potentials
Share
Study at a glance
- Area
- QEEG, Methods & Mechanisms
- Study type
- Controlled study
- Authors
- Tallgren P, Vanhatalo S, Kaila K, Voipio J
- Journal
- Clinical Neurophysiology
- Year
- 2005
- Protocol
- bench testing of EEG electrodes and gels
Key findings
- Reusable sintered Ag/AgCl electrodes gave the best results for slow-potential recording
- All six disposable Ag/AgCl electrode models had suitable electrical properties
- Silver, tin, gold, platinum and stainless steel electrodes showed varying polarization, baseline drift, noise, high resistance or wear-related changes
- Seven of nine gels contained substantial chloride, which DC stability requires, while the exact chloride concentration mattered little
A laboratory comparison of electrodes and conductive gels for recording very slow EEG potentials, which requires direct current (DC) stable recording. The authors tested six reusable electrode types, six disposable silver-silver chloride (Ag/AgCl) models and nine gels or pastes for polarization, stability and low-frequency noise. Reusable sintered Ag/AgCl electrodes performed best, and the disposable Ag/AgCl models were also adequate. The authors conclude that DC-coupled EEG depends heavily on the choice of electrode and gel, and they offer criteria for selecting them.