Study at a glance

Area
QEEG, Methods & Mechanisms
Study type
Controlled study
Authors
Tallgren P, Vanhatalo S, Kaila K, Voipio J
Journal
Clinical Neurophysiology
Year
2005
Protocol
bench testing of EEG electrodes and gels

Key findings

  • Reusable sintered Ag/AgCl electrodes gave the best results for slow-potential recording
  • All six disposable Ag/AgCl electrode models had suitable electrical properties
  • Silver, tin, gold, platinum and stainless steel electrodes showed varying polarization, baseline drift, noise, high resistance or wear-related changes
  • Seven of nine gels contained substantial chloride, which DC stability requires, while the exact chloride concentration mattered little

A laboratory comparison of electrodes and conductive gels for recording very slow EEG potentials, which requires direct current (DC) stable recording. The authors tested six reusable electrode types, six disposable silver-silver chloride (Ag/AgCl) models and nine gels or pastes for polarization, stability and low-frequency noise. Reusable sintered Ag/AgCl electrodes performed best, and the disposable Ag/AgCl models were also adequate. The authors conclude that DC-coupled EEG depends heavily on the choice of electrode and gel, and they offer criteria for selecting them.